Responsiveness of 2 Procedures for Measurement of Temporal and Spatial Gait Parameters in Older Adults

Received 20 July 2009; accepted 17 February 2010.

Objective

To determine the responsiveness of the GAITRite system and a stopwatch-footfall count technique for measurement of walking speed, cadence, and stride length during comfortable and fast-paced walking.

Design

Criterion standard.

Setting

Research laboratory in a physical therapy education program.

Participants

Twenty-four healthy volunteers (13 men, 11 women; mean age 74.5 years) without lower extremity injury or history of falls.

Interventions

Participants walked across a GAITRite mat with embedded pressure sensors at their self-selected comfortable and fast walking speeds. Simultaneously, an examiner, using a stopwatch, recorded the elapsed time necessary to cross the mat and counted the number of complete footfalls.

Main Outcome Measure(s)

Walking speed, cadence, and stride length were compared between the GAITRite system and the stopwatch-footfall count technique for both comfortable and fast walking speeds. Responsiveness values for each procedure were described by the 95% minimal detectable change (MDC).

Results

During comfortable self-paced walking, MDC values for the stopwatch-footfall count technique ranged from 10% to 65% greater than those obtained for the GAITRite system. During fast self-paced walking MDC values for the stopwatch-footfall count technique ranged from 26% to 65% larger than those measured by the GAITRite system for the temporal and spatial gait performance parameters.

Conclusions

When measured by the GAITRite system, the 95% MDC values for temporal and spatial gait parameters of older community-dwelling adults were more responsive to change than those obtained by the stopwatch-footfall technique. Clinicians should recognize that self-selected walking speed, cadence, and stride length when obtained by an instrumented walkway must be equal to or exceed 12.6 cm/s, 8.4 steps/min, or 7 cm, respectively, for the change to be considered real change and not from measurement error.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 
  •  Disclosure: nothing to disclose
  •  Disclosure: nothing to disclose
  • § Disclosure: nothing to disclose
  •  Disclosure: nothing to disclose
  •  Disclosure: nothing to disclose
  • # Disclosure: nothing to disclose

 No commercial party having a direct financial interest in the results of the research supporting this article will confer a benefit upon the authors or upon any organization with which the authors are associated.

 Funding source: Mayo Program in Physical Therapy.

 Disclosure Key can be found on the Table of Contents and at www.pmrjournal.org

PII: S1934-1482(10)00133-4

doi:10.1016/j.pmrj.2010.02.008

PM&R
Volume 2, Issue 6 , Pages 537-543, June 2010